Dear lab colleagues, are you struggling with choosing the specifications for Puredex® G-25M Desalting Prepacked Column? 1mL, 4.7 mL, 5mL, 20mL... Faced with these options, how to make the best choice?
Today, we let real desalting chromatogram data and column specifications speak for themselves—revealing the science behind selecting the ideal Puredex® G-25M prepacked caolumn.
First, let's look at the size differences between the four empty columns:
Specification | Column Dimensions (Diameter × Height) | Bed Volume | Design Features |
1mL | 7.7×25mm | ~1 mL | Short and narrow, for rapid screening tests |
4.7mL | 7.7×100mm | ~4.7 mL | Tall and narrow, prioritizes resolution |
5mL | 16×25mm | ~5 mL | Short and wide, prioritizes throughput |
20mL (16/100) | 16×100mm | ~20 mL | Standard column, balanced design |
Table 1: Specification dimensions of empty columns
Comparatively, the design of 1mL 7.7 mm narrow diameter and short column bed height determines its rapid screening characteristics; The design of 7.7 mm narrow diameter and long column bed height of 4.7 mL determines its priority on resolution; Both 5mL and 20mL are designed with a wide diameter of 16mm, which determines their priority on throughput.

Figure 1: Puredex® G-25M Prepacked Column
Different volumes Puredex® G-25M Desalting Comparison:
Resin: Puredex® G-25M
Chromatography Column: 1mL, 4.7 mL, 5mL, 20mL
Sample: BSA (50 mM PB, 500 mM NaCl, pH 7.0)
Sample Volume: 5% column volume
Flow Rate: 100 cm/h
Buffer: 50 mM PB, pH 7.0


Figure 2: Puredex® G-25M desalting chromatogram
From the analysis of the desalting chromatogram, it was found that:
1mL (7.7×25mm):
Advantages: Narrow and short type can quickly screen resins
Limitations: Low column bed height, incomplete sample desalination, and actual separation effect is not optimal
4.7mL (7.7×100mm):
Advantages: Narrow diameter, 100mm column bed height, highest theoretical resolution
Limitations: Too narrow diameter, limited flow rate, easy to generate high pressure, not suitable for large volume samples
5mL (16×25mm):
Golden Combination: 16mm diameter provides excellent flow rate, 25mm height optimizes separation time
Chromatogram Shows: High separation efficiency, ability of rapid sample desalting.
20mL (16×100mm):
Process-level Design: 16mm diameter ensures flow rate, 100mm column height provides large sample loading space
Chromatogram Shows: Sharp and symmetrical peak shape, high separation efficiency, and minimal sample dilution effect
First choice 5mL (16 × 25 mm)-The undisputed “workhorse" for daily experiments:
Daily Post Protein Purification Buffer Change
Desalination treatment of precious samples such as antibodies and enzymes
Routine experiments that need to be completed quickly and guaranteed recovery
Key Advantages: Short column bed (25mm) reduces sample residence time in the column, reduces sample dilution, and the recovery rate is usually > 95%
Upgraded 20 mL (16 × 100 mm)-the "expert” for large volume processing:
Preliminary treatment of crude samples such as cell culture supernatant and fermentation broth
Stability validation during process development stage, which requires multiple repeated experiments
Direct handling of high concentration samples
Key advantage: 100mm column height provides sufficient separation distance to ensure resolution even for large samples
Why Is Diameter Change So Important?
7.7 mm Narrow Column: Suitable for analytical-grade applications but sensitive to sample viscosity
16mm Wide Column: Reduces column pressure, improve processing capacity, and is more suitable for preparation-grade needs
How Does Column Height Affect Separation?
25mm Short Column: Quick separation, suitable for simple desalination (buffer change)
100mm Long Column: Provides more theoretical plates, suitable for complex samples or higher purity needs
"Chunky" Advantage of 5 mL (16 × 25 mm)
Low Pressure Operation: Increased diameter, lower column pressure at same flow rate
Quick Separation: 25mm column height means shorter runtime
When 5 mL (16 × 25 mm) Is Used:
Recommended operating flow rate: 1-10 mL/min
Recommended sample volume: 0.1-1.5 mL
Regeneration recommendation: Can be reused 5-10 times to maintain good performance
When 20 mL (16 × 100 mm) is used:
Optimal flow rate: 3-14 mL/min
Recommended sample volume: ≤ 6 mL
Special application: Suitable for handling crude samples containing tiny particles, not easy to block
When using pre-packed columns of different specifications, please note:
For all specifications, avoid column bed drying.
Viscosity significantly affects performance. Reduce flow rate for high-viscosity samples.
It is recommended to optimize the conditions with small size first, and then enlarge them to large size in parallel.
The specification differences of Puredex® G-25M desalting prepacked columns are not merely changes in volume numbers, but rather a reflection of the scientific design of column dimensions. Through the clever combination of diameter and height, the 5mL (16 × 25 mm) size achieves a golden balance of flux and resolution, while the 20mL (16 × 100 mm) size provides a reliable solution for large-volume sample processing.
Understanding the size science behind these can make your experimental design more accurate and efficient!
Cat. No.
Bulk Resin | ||
Product Name | Cat.No. | Pack Size |
Puredex® G-25M | 1050-2140 | 25 g |
1050-2148 | 50 g | |
1050-2141 | 100 g | |
1050-2142 | 500 g | |
1050-2143 | 1 kg | |
1050-2144 | 5 kg | |
Pre-packed Volume | ||
Product Name | Cat.No. | Pack Size |
Chrom-Trap® Puredex® G-25M 5 mL | 2050-2143 | 1×5 mL |
Puredex® G-25M 16/100 | 2050-2144 | 16/100 |
Link With Us Now!